Viewing Study NCT05542732


Ignite Creation Date: 2025-12-24 @ 9:25 PM
Ignite Modification Date: 2025-12-27 @ 7:50 AM
Study NCT ID: NCT05542732
Status: COMPLETED
Last Update Posted: 2022-09-15
First Post: 2022-09-09
Is NOT Gene Therapy: True
Has Adverse Events: False

Brief Title: Accuracy of CardioWatch 287-2 During Profound Hypoxia
Sponsor: Corsano Health B.V.
Organization:

Study Overview

Official Title: Accuracy of CardioWatch 287-2 During Profound Hypoxia
Status: COMPLETED
Status Verified Date: 2022-09
Last Known Status: None
Delayed Posting: No
If Stopped, Why?: Not Stopped
Has Expanded Access: False
If Expanded Access, NCT#: N/A
Has Expanded Access, NCT# Status: N/A
Acronym: ACW2
Brief Summary: In an individual's SpO2 range from 100 to 73 %, this study calibrates and evaluates the accuracy of SpO2 measurement by CW2 in comparison to reference pulse oximeter (Nellcor PM10) or CO-oximetry, Pulse rate (PR) and respiratory rate (RR) were calibrated based on the same Nellcor equipment and the frequency of end-tidal CO2 respectively.
Detailed Description: The study is designed in accordance with Medical electrical equipment - Part 2-61: Particular requirements for basic safety and essential performance of pulse oximeter equipment ISO 80601-2-61:2017, Corrected version 2018-02 and ISO 80601-2-55.

SpO2: Testing to confirm the accuracy of the SpO2 monitoring according to ISO 80601-2-61 Clause 201.12.1.101 201.12.1.101

RR Testing: Testing to confirm the accuracy of the respiration rate monitoring according to ISO 80601-2-55. Volunteers respirated spontaneously, no efforts were made to include all ranges of respiration during the study.

PR Testing: Testing to confirm the accuracy of the PR monitoring according to ISO 80601-2-61 Clause 201.12.1.104. No efforts were made to influence PRs of volunteers.

An accelerometer and PPG filters are built in CW2 in order to form the basis to exclude non-eligible test results.

Study Oversight

Has Oversight DMC: False
Is a FDA Regulated Drug?: False
Is a FDA Regulated Device?: False
Is an Unapproved Device?: None
Is a PPSD?: None
Is a US Export?: None
Is an FDA AA801 Violation?: