Viewing Study NCT06078956



Ignite Creation Date: 2024-05-06 @ 7:37 PM
Last Modification Date: 2024-10-26 @ 3:10 PM
Study NCT ID: NCT06078956
Status: RECRUITING
Last Update Posted: 2023-10-12
First Post: 2023-02-14

Brief Title: ABL90 Flex Plus Method Comparison Study_Adults
Sponsor: Radiometer Medical ApS
Organization: Radiometer Medical ApS

Study Overview

Official Title: ABL90 FLEX PLUS Method Comparison Study for pH pCO2 pO2 cCa2 cCl- cK cNa cGlu cLac ctHb sO2 FO2Hb FMetHb FCOHb and FHHb in Whole Blood in Adult Population
Status: RECRUITING
Status Verified Date: 2023-10
Last Known Status: None
Delayed Posting: No
If Stopped, Why?: Not Stopped
Has Expanded Access: False
If Expanded Access, NCT#: N/A
Has Expanded Access, NCT# Status: N/A
Acronym: None
Brief Summary: The goal of this clinical trial is to validate performance claims for method comparison for the ABL90 FLEX PLUS for pH pCO2 pO2 cCa2 cCl- cK cNa cGlu cLac ctHb sO2 FO2Hb FMetHb FCOHb and FHHb in heparinized whole blood in a Point of Care POC setting
Detailed Description: The main question it aims to answer is

To quantify the relationship using comparison measurement from the same sample between ABL90 FLEX PLUS and the predicate device within the reportable range for all 15 parameters in heparinized arterial and venous patient whole blood in syringe measuring mode S65 SP65 and heparinized capillary whole blood samples in capillary measuring mode C65

Study Oversight

Has Oversight DMC: None
Is a FDA Regulated Drug?: False
Is a FDA Regulated Device?: True
Is an Unapproved Device?: None
Is a PPSD?: None
Is a US Export?: None
Is an FDA AA801 Violation?: None