Viewing Study NCT05563701



Ignite Creation Date: 2024-05-06 @ 6:09 PM
Last Modification Date: 2024-10-26 @ 2:42 PM
Study NCT ID: NCT05563701
Status: COMPLETED
Last Update Posted: 2022-10-03
First Post: 2022-09-22

Brief Title: Evaluation of the LVivo Image Quality Scoring IQS
Sponsor: DiA Imaging Analysis Ltd
Organization: DiA Imaging Analysis Ltd

Study Overview

Official Title: Evaluation of the LVivo Image Quality Scoring IQS
Status: COMPLETED
Status Verified Date: 2022-09
Last Known Status: None
Delayed Posting: No
If Stopped, Why?: Not Stopped
Has Expanded Access: False
If Expanded Access, NCT#: N/A
Has Expanded Access, NCT# Status: N/A
Acronym: None
Brief Summary: The study includes two parts

Part 1

100 examinations of patient referred for echo evaluation containing clips that were acquired from the 4CH apical view were collected retrospectively Each study includes 10 clips that represent typical user scanning errors
Offline evaluation of the system by batch processing shall be performed by comparing the system output to preliminary quality tagging by experienced sonographers

Part 2

Live scans of apical 4CH clips of patients with indication for POCUS examination will be performed by POC physicians
LVivo IQS shall be used on Lumify during the scan for patients that meet inclusion criteria until 50 exams will be collected 3 sec of each scan shall be saved and Image quality score IQS shall be documented
Saved scans shall be reviewed by an expert physician to determine whether they are clinically interpretable
Detailed Description: Part 1 Total of 424 examination were collected from two medical centers Participants that were referred to routine echo examinations and signed an informed consent were enrolled to the study The image acquisition was done using ultrasound devices available in the medical centers In addition Lumify ultrasound device was supplied by the sponsor

Each examination contained 10 clips each representing a different scanning scenario where the image acquisition contained transitions between common user acquisition error to correctly acquired images

The examinations in DICOM format and echo reports were collected after anonymization the patient details were documented in a study log that was stored in the medical center

The collected clips were tagged for image quality by experienced sonographers The distribution of the collected data according to ultrasound devices age gender and LV function was analyzed

The data was randomly split into training and clinical validation sets such that the proportion of data according to data distribution was maintained and the data set for the clinical evaluation will contain 100 patient examinations

The LVivo IQS algorithm will be applied to the clinical validation set automatically by batch processing The image quality score provided by the LVivo IQS will be compared to the image quality tagging by the sonographers

Part 2 The trial is designed as a prospective multicenter study enrolling 50 patients at two medical centers In each medical center Lumify ultrasound devices on which the LVivo IQS demo software will be installed will be provided for image acquisition The image acquisition will be done by POC medical doctors at different stages of their internship that received POCUS training during their medical studies and are using ultrasound devices in accordance with their routine workflow The scans will be done in different POC units including ER ICU and internal departments

Study Oversight

Has Oversight DMC: None
Is a FDA Regulated Drug?: None
Is a FDA Regulated Device?: None
Is an Unapproved Device?: None
Is a PPSD?: None
Is a US Export?: None
Is an FDA AA801 Violation?: None