Brief Title: Evaluation of the esCCO Non-invasive Cardiac Output Measurement Device in Pregnancy
Sponsor: Kings College Hospital NHS Trust
Organization: Kings College Hospital NHS Trust

Study Overview

Official Title: Evaluation of the esCCO Non-invasive Cardiac Output Measurement Device in Pregnancy
Status: UNKNOWN
Status Verified Date: 2014-09
Last Known Status: NOT_YET_RECRUITING
Delayed Posting: No
If Stopped, Why?: Not Stopped
Has Expanded Access: False
If Expanded Access, NCT#: N/A
Has Expanded Access, NCT# Status: N/A
Acronym: esCCO
Brief Summary: Cardiac output CO is the volume of blood ejected from the heart and is a product of stroke volume SV vs heart rate HR It is closely related to Blood Pressure BP by the formula CO Mean Arterial Pressure MAP Systemic Vascular Resistance SVR Cardiac output monitoring is well established and validated for guiding fluid administration and resuscitation in intensive care and perioperatively in the non-pregnant population Available CO monitoring methods may risk maternal health or fetal wellbeing or are unvalidated in the pregnant population Invasive and semi-invasive methods of measuring CO such as the pulmonary artery catheter the gold standard require invasive arterial or central access negating their use in all but a select group of labouring women and have potentially serious risks attached to their usage Echocardiography is a well-established and validated technique requiring no invasive access but the requirement for an experienced operator limits routine usage It uses an external ultrasound probe to scan the heart

Several novel non-invasive cardiac output monitors have recently come to the market including the estimated cardiac output esCCO monitor Nihon Kohden which uses Pulse Wave Transit Time PWTT to estimate cardiac output It requires 3-lead Electrocardiography and pulse oximetry alone which is part of the routine monitoring for high risk patients on the labour ward

The investigators research aims to evaluate the accuracy and precision of the esCCO in pregnant women and subsequently assess its utility during medical interventions such as epidural analgesia or caesarean section The initial validation will take place in an antenatal clinic where women are assessed using Doppler echocardiography The investigators will then compare the echocardiography results vs the esCCO results
Detailed Description: The cardiac output CO is the volume of blood ejected by the heart each time it beats It is the product of stroke volume SV and heart rate HR and is expressed in litres per minute It gives a measure of the performance of the heart as a pump The normal resting CO of a 70kg person is approximately 5 to 6 litresmin due to a resting HR of 70-80 bpm multiplied by an average SV of 70-80ml Different organs in the body receive differing proportions of the CO according to their demand CO increases significantly in pregnancy due to the growing uterus fetus and placenta and its actual value can be affected by patient position and medical interventions such as epidural analgesia

Cardiovascular changes in pregnancy

During pregnancy the cardiovascular system changes to adapt to the demands of the uteroplacental unit CO starts to increase from the first trimester by 35-40 to a 50 rise by the end of the second trimester It remains at this level during the third trimester 1 The increase in CO occurs due to an increase in HR by up to 25 and SV by up to 30 There is a further rise in CO during labour and delivery which is transient

CO also depends upon patient position with a 135 increase in the left lateral position from supine The increase in CO in the left lateral position is the basis for the left lateral tilt or wedge applied to the pregnant patient when supine This reduces compression of the large abdominal blood vessels aortocaval compression by the gravid pregnant uterus and allows the CO to be maintained

The proportion of the CO distributed to the uterus kidneys and skin is greater in pregnancy compared to the non-pregnant state Uterine blood flow is approximately 500 to 700 mlmin which corresponds to 10-12 of the CO 1 More than 80 of this CO perfuses the placenta

Lumbar epidurals in labour

Lumbar epidurals are used commonly for analgesia in labour as well as for anaesthesia for operative interventions such as lower segment Caesarean section LSCS and instrumental delivery It is well known that epidural anaesthesia can cause haemodynamic changes namely peripheral vasodilatation and a reduction in systemic vascular resistance SVR due to blockade of the sympathetic nervous system which controls many autonomic functions including blood pressure and heart rate 1 This can lead to a reduction in blood pressure if the CO is maintained since the three are related by the formula mean arterial pressure MAP CO x SVR If the CO drops for example due to hypovolaemia decreased amount of fluid in the circulation or postural changes then there will be a greater decrease in the blood pressure

Valsalva manoeuvre Originally described in the 17th century the Valsalva manoeuvre is performed by attempted exhalation against resistance usually a closed airway It is used routinely to equalize ambient pressure in the inner ear and sinuses ie during air travel or scuba diving Medically it is used as a test of cardiac function and the autonomic nervous system which controls the hearts response to stimuli

The standardised Valsalva manoeuvre requires a subject to be placed in a semirecumbent position and produce an expiratory pressure of 40mmHg for 10 seconds2

Methods for measuring cardiac output There are several methods for measuring cardiac output but their use in pregnancy may be limited by risks to mother or the effects of the technique on the developing fetus

Invasive methods - The gold standard CO monitor is the pulmonary artery catheter This measures CO using the temperature change of an injectate the thermodilution principle as blood is passing through the cardiac chambers It is accurate and allows measurement of pulmonary artery pressures but the obvious drawback is the invasive nature of the device coupled with the risks of injury to the heart and pulmonary vessels Due to its invasive nature this is used almost exclusively in patients who are severely unwell necessitating central catheterisation Even so there is currently no evidence from randomised controlled trials supporting the use of the pulmonary artery catheters 2 Its use is not possible in normal pregnancies as it would be unacceptable to subject patients to the risks of central catheterisation Other invasive techniques such as angiography or conductance catheters cause radiation exposure to the developing fetus

Semi-invasive methods include LiDCO and PiCCO The LiDCO monitor uses transpulmonary lithium dilution via a central line catheter and arterial sampling line for calibration It then uses arterial waveform analysis to provide continuous CO measurements PiCCO uses transpulmonary thermodilution again via a central line and arterial sampling line The disadvantages of these are that lithium dilution should be avoided in the first trimester and normal pregnancies do not require central or arterial lines This negates their use in the normal pregnant population However newer LiDCO products such as the LiDCOrapid have been used in the literature 3

Non-invasive methods for cardiac output measurement have been the cornerstone of studying maternal cardiac physiology in normal and pathological pregnancies in the last 20 years since the introduction of Doppler echocardiography around 1985 4 Echocardiographic techniques aim to estimate the SV which can then be multiplied by the HR to obtain a CO value This can be done using 2D 3D or Doppler echocardiography 5 The obvious advantage is that no vascular access or calibration is needed but an experienced operator is required making routine use difficult

Novel non-invasive cardiac output monitors have since appeared on the market such as the noninvasive cardiac output monitor NICOM the ultrasonic cardiac output monitor USCOM and the estimated continuous cardiac output esCCO monitor

The esCCO monitor The esCCO monitor is manufactured by Nihon Kohden6 and uses 3 lead electrocardiogram and a pulse oximeter oxygen saturation probe to estimate CO

This method of measuring CO was shown to correlate well with the trend of CO measured using the pulmonary artery catheterisation method7 8 It has also been shown to correlate well with the echo Doppler during exercise9 Bataille et al 2012 compared the esCCO monitor with transthoracic echocardiography for CO measurement in intensive care patients10 The authors deemed the esCCO to be clinically unacceptable in critically ill patients although there were questions about the validity of this study11 12 This is primarily because of extensive vasoconstriction and peripheral shut down with the critically ill septic patients Use of this device in obstetrics is acceptable

The esCCO monitor uses non-invasive blood pressure NIBP three-lead ECG and pulse oximetry SpO2 These modalities are part of routine clinical monitoring which makes this monitor ideal for measuring CO in the normal pregnant population Patients admitted to the labour ward often require monitoring with at least NIBP and SpO2 hence only the three-lead ECG is extra By validating this monitor CO measurement across a variety of clinical scenarios in the pregnant population will be possible