Detailed Description:
Medical background and current standard of care:
Osteoporosis is defined as a systemic skeletal disease characterized by low mineral bone mass and microarchitectural deterioration of bone tissue (Consensus Conference, 1994). This increased bone fragility is associated with susceptibility to fractures, which have substantial economic and societal burdens. Osteoporosis is routinely diagnosed through bone density assessments; however, there is a shortage of accurate means to characterize bone microstructure for diagnostic purposes. The POROUS R4C ultrasound device is capable of providing such information in detail.
The current standard of care (SoC) physical bone parameter for examining bone health is areal BMD (aBMD), typically measured by dual-energy X-ray absorptiometry (DXA) at the hip or spine. Scaling is given to score the patient's aBMD in comparison to a reference of a healthy bone. The resulting T Score compares the patient's aBMD to the peak aBMD of young adults of the same sex, while the Z score compares it to the average aBMD of the same age and sex. According to the operational WHO definition (1994), a T Score of -1.0 or above is considered normal, while scores between -1.0 and -2.5 indicate osteopenia, and scores of -2.5 or lower indicate osteoporosis.
The rationale for the clinical investigation:
Quantitative ultrasound (QUS) devices have several substantial advantages over X-ray-based technologies: a better safety profile (non-ionizing radiation), cost-efficiency, usability, and portability. These advantages can result in improved accessibility at point-of-care. The major challenge is, nonetheless, clinical performance. QUS devices that assess bone health have demonstrated varying performance in estimating surrogate aBMD values. aBMD estimation is calculated by empirical correlation with DXA reference measurements using a combination of ultrasound parameters and anthropometric information. We follow these clinical performance standards for aBMD estimation using the investigational POROUS R4C ultrasound device.
The investigational device:
The POROUS R4C is a non-invasive QUS device for measuring and quantifying microstructural, acoustic, and viscoelastic properties in cortical bone (e.g., in the tibia of the lower limb). The device can be used to acquire single-channel raw data in a pulse-echo mode and perform innovative signal processing and imaging techniques to determine physical bone biomarkers. Examples of physical bone biomarkers are cortical thickness and the speed of sound in the radial direction using refraction-corrected image reconstruction and frequency-dependent attenuation and pore-size distribution using spectral backscatter analysis. The measurement data are recorded by the acquisition software, and the post-processing analysis is done using separate algorithms. In pre-clinical and clinical studies (completed, ongoing, or planned), the effectiveness of spectral ultrasound backscatter analysis and refraction-corrected image reconstruction methods are being evaluated to assess bone quality, health, and biomechanical competence. In first-in-human pilot studies, the POROUS technology has demonstrated the feasibility of measuring cortical bone properties and estimating aBMD in smaller cohorts of participants.
The overall purpose of the clinical investigation:
In this confirmatory clinical investigation, data will be collected to characterize cortical bone properties and calculate surrogate aBMD values for young and middle-aged (21 to 55 years) men and women. In a separate confirmatory clinical investigation (POROUS-preFX, NCT06567054), data will be collected to characterize cortical properties and calculate surrogate aBMD values in older men and women (56 to 85 years). Together, the combined data will represent a broader range of estimated surrogate aBMD values and generate a POROUS T score for diagnostic purposes.
Clinical investigation design:
This is a cross-sectional, single-cohort, age- and sex-stratified study of participants aged 21 to 55 years. The rationale for selecting this age range is to gather data for this age group, while data from older participants are gathered in an independent clinical investigation (POROUS-preFX, NCT06567054).
Participants will be enrolled into different groups based on their age (consisting of five-year bands) and sex (males or females). All participants will undergo measurements using the investigational POROUS R4C ultrasound device and the comparator DXA device. Prior knowledge of aBMD values, whether measured by DXA or another type of bone densitometer, will not be recorded to avoid selection bias. The final analysis of study results will incorporate measured POROUS parameters (e.g., cortical thickness, cortical bone pore diameter distribution, cortical bone frequency-dependent ultrasound attenuation, cortical bone backscatter coefficient, speed of sound in the radial direction), anthropometric information (weight, height, BMI, sex), and age will be used to create a multivariate partial least squares (PLS) model to predict DXA-derived aBMD. This prediction model will be similar to the one described in the pilot studies (Armbrecht et al., JBMR Plus, 2021; Dehnen et al., Front Endocrinol, 2024). However, the formula will be re-developed based on the results of this study. The predictions will be made with DXA-derived aBMD values of the femur neck, total proximal femur, and lumbar spine collected in this study. Cross-calibration of DXA devices from different manufacturers will be done if necessary.
The participants will also receive a questionnaire regarding their medical history and any medications that may impact bone health.
Primary objective:
• To assess the performance of the POROUS R4C ultrasound device in estimating surrogate aBMD values (total femur).
Secondary objectives:
* To assess the performance of the POROUS R4C ultrasound device in estimating surrogate aBMD values (femoral neck and lumbar spine).
* To assess the safety of the POROUS R4C ultrasound device by monitoring adverse events affecting participants or healthcare professionals using the device.
Exploratory objectives:
* To generate T-score and Z-score scales (total femur, femoral neck, and lumbar spine) for the POROUS R4C ultrasound device and assess their performance.
* To collect data on cortical bone properties measured by POROUS R4C ultrasound device in young and middle-aged men and women.
* To assess the reproducibility of direct measurements by the POROUS R4C
Participants:
The clinical investigation will include 350 female and male participants. Prior knowledge of bone health or aBMD values will not be considered an enrolment criterion. All participants will undergo measurement using the investigational POROUS R4C ultrasound device at the midshaft tibia and a comparator DXA device at the total femur, femoral neck, and lumbar spine. The participants are stratified according to sex and age stratification into five-year bands:
* Age group 21-25, males/females = 25 per group, total = 50.
* Age group 26-30, males/females = 25 per group, total = 50.
* Age group 31-35, males/females = 25 per group, total = 50.
* Age group 36-40, males/females = 25 per group, total = 50.
* Age group 41-45, males/females = 25 per group, total = 50.
* Age group 46-50, males/females = 25 per group, total = 50.
* Age group 31-55, males/females = 25 per group, total = 50.
Screened and eligible individuals will be enrolled in the investigation until the necessary sample size for his/her corresponding group (based on age and sex) has been reached (see Table above). To avoid over-recruiting, once the necessary sample size for one group has been reached, no further individuals with matching age band or sex will be enrolled in the investigation.
Duration of the clinical investigation:
The planned overall clinical investigation is expected to last up to 12 months, from the first participant's enrolment until the last participant's clinical assessment. However, if the last participant's assessment is completed before the 12-month mark, the Sponsor may choose to conclude the investigation earlier.
Clinical procedures:
The following clinical procedures are performed during the investigation:
Screening:
* Inclusion/exclusion criteria
* Enrolment into corresponding age and sex groups.
* Informed consent
Clinical assessments:
* Measuring with the POROUS R4C ultrasound device at the central antero-medial tibia region
* Scanning with DXA (aBMD of the lumbar spine and proximal femur).
* Assessment of medical conditions and concomitant medication using a questionnaire.
* Recording of adverse events.