Viewing Study NCT05962905


Ignite Creation Date: 2025-12-25 @ 3:32 AM
Ignite Modification Date: 2025-12-26 @ 2:14 AM
Study NCT ID: NCT05962905
Status: COMPLETED
Last Update Posted: 2025-05-09
First Post: 2023-07-13
Is NOT Gene Therapy: False
Has Adverse Events: False

Brief Title: Accuracy Performance Verification of the INVOS™ System in Pediatrics
Sponsor: Medtronic - MITG
Organization:

Study Overview

Official Title: Medtronic Accuracy Performance Verification of the INVOS™ System in the Pediatric Cardiac Catheterization Laboratory (INCATH)
Status: COMPLETED
Status Verified Date: 2025-05
Last Known Status: None
Delayed Posting: No
If Stopped, Why?: Not Stopped
Has Expanded Access: False
If Expanded Access, NCT#: N/A
Has Expanded Access, NCT# Status: N/A
Acronym: INCATH
Brief Summary: Prospective, observational, multi-center trial designed to capture calibration and performance evaluation data with the INVOS™ system.
Detailed Description: The purpose of this study is to verify INVOS™ cerebral oximetry system performance in pediatric and infant patients bycomparing regional oxygen saturation (rSO2) values to reference blood oxygen saturation measurements obtained throughinternal jugular vein and arterial blood convenience sampling according to International Organization on Standards (ISO) 80601-2-85:2021.

Study Oversight

Has Oversight DMC: True
Is a FDA Regulated Drug?: False
Is a FDA Regulated Device?: True
Is an Unapproved Device?: None
Is a PPSD?: None
Is a US Export?: True
Is an FDA AA801 Violation?: