Viewing Study NCT01727505


Ignite Creation Date: 2025-12-25 @ 3:26 AM
Ignite Modification Date: 2026-02-28 @ 6:33 PM
Study NCT ID: NCT01727505
Status: COMPLETED
Last Update Posted: 2017-03-01
First Post: 2012-11-08
Is NOT Gene Therapy: True
Has Adverse Events: True

Brief Title: Volume Guarantee Ventilation in Preterm Infants With Frequent Episodes of Hypoxemia
Sponsor: University of Miami
Organization:

Study Overview

Official Title: Volume Guarantee Ventilation in Preterm Infants With Frequent Episodes of Hypoxemia
Status: COMPLETED
Status Verified Date: 2017-01
Last Known Status: None
Delayed Posting: No
If Stopped, Why?: Not Stopped
Has Expanded Access: False
If Expanded Access, NCT#: N/A
Has Expanded Access, NCT# Status: N/A
Acronym: None
Brief Summary: Preterm infants undergoing mechanical ventilation often present with respiratory instability which leads to fluctuations in ventilation and oxygenation. In conventional modes of mechanical ventilation used in these patients, the ventilator delivers breaths at a set rate and a constant peak inspiratory pressure.

We have demonstrated that spontaneous episodes of hypoxemia in mechanically ventilated preterm infants are frequently triggered by a loss in end-expiratory lung volume followed by decrease in tidal volume. This is due to a reduction in respiratory system compliance and increase in airway resistance.

Volume Guarantee ventilation is a mode of ventilation in which ventilator measures the exhaled tidal volume of each ventilator breath and automatically adjusts the peak inspiratory pressure to deliver the set tidal volume. We have previously shown that Volume Guarantee reduced the severity and duration of hypoxemia episodes during a period of 2 hours in preterm infants The study hypothesis is that Volume Guarantee ventilation will decrease the duration of episodes of hypoxemia when compared to conventional mechanical ventilation during routine clinical conditions and over longer periods of time.

The objective of the study is to evaluate the effects of Volume Guarantee on the duration, severity and frequency of episodes of hypoxemia in mechanically ventilated preterm infants who present with frequent hypoxemia episodes in comparison to conventional ventilation over two periods of 24 hours each and under routine clinical conditions.
Detailed Description: None