The Description Module contains narrative descriptions of the clinical trial, including a brief summary and detailed description. These descriptions provide important information about the study's purpose, methodology, and key details in language accessible to both researchers and the general public.

Description Module path is as follows:

Study -> Protocol Section -> Description Module

Description Module


Ignite Creation Date: 2025-12-24 @ 10:51 PM
Ignite Modification Date: 2025-12-24 @ 10:51 PM
NCT ID: NCT04360369
Brief Summary: Determine if the Tono-Vera Tonometer accurately measures intraocular pressure (IOP). The hypothesis of this test is to confirm the Tono-Vera Tonometer is equivalent to the Goldmann Applanation Tonometer (within +/- 5.0 mmHg).
Detailed Description: The Tono-Vera Tonometer (Reichert, Inc.) is a portable rebound tonometer intended to measure IOP. Subjects will undergo a standard ophthalmologic examination. Subjects will then be measured with four tonometers: Goldmann Applanation Tonometer (C.S.O. SRL), Ocular Response Analyzer G3 (Reichert, Inc.), iCare ic100 tonometer (iCare Finland Oy), and Tono-Vera Tonometer.