The Description Module contains narrative descriptions of the clinical trial, including a brief summary and detailed description. These descriptions provide important information about the study's purpose, methodology, and key details in language accessible to both researchers and the general public.

Description Module path is as follows:

Study -> Protocol Section -> Description Module

Description Module


Ignite Creation Date: 2025-12-24 @ 9:25 PM
Ignite Modification Date: 2025-12-24 @ 9:25 PM
NCT ID: NCT05542732
Brief Summary: In an individual's SpO2 range from 100 to 73 %, this study calibrates and evaluates the accuracy of SpO2 measurement by CW2 in comparison to reference pulse oximeter (Nellcor PM10) or CO-oximetry, Pulse rate (PR) and respiratory rate (RR) were calibrated based on the same Nellcor equipment and the frequency of end-tidal CO2 respectively.
Detailed Description: The study is designed in accordance with Medical electrical equipment - Part 2-61: Particular requirements for basic safety and essential performance of pulse oximeter equipment ISO 80601-2-61:2017, Corrected version 2018-02 and ISO 80601-2-55. SpO2: Testing to confirm the accuracy of the SpO2 monitoring according to ISO 80601-2-61 Clause 201.12.1.101 201.12.1.101 RR Testing: Testing to confirm the accuracy of the respiration rate monitoring according to ISO 80601-2-55. Volunteers respirated spontaneously, no efforts were made to include all ranges of respiration during the study. PR Testing: Testing to confirm the accuracy of the PR monitoring according to ISO 80601-2-61 Clause 201.12.1.104. No efforts were made to influence PRs of volunteers. An accelerometer and PPG filters are built in CW2 in order to form the basis to exclude non-eligible test results.
Study: NCT05542732
Study Brief:
Protocol Section: NCT05542732