Description Module

Description Module

The Description Module contains narrative descriptions of the clinical trial, including a brief summary and detailed description. These descriptions provide important information about the study's purpose, methodology, and key details in language accessible to both researchers and the general public.

Description Module path is as follows:

Study -> Protocol Section -> Description Module

Description Module


Ignite Creation Date: 2025-12-24 @ 2:55 PM
Ignite Modification Date: 2025-12-24 @ 2:55 PM
NCT ID: NCT02094859
Brief Summary: The primary purpose of this Study is evaluation of the safety and performance of the GlucoClear System (System). The primary safety objective will be assessed by: • Evaluation of any serious adverse device effects upon Sensor insertion through Sensor removal and a follow-up assessment one week (7 ± 3 days) after Sensor removal. The primary performance objective will be evaluated using the following criteria: 15/20% Performance: Blood glucose measurements from the System must meet the ISO 15197 15/20% accuracy criteria (at least 95% of the matched pairs within ± 15 mg/dL of Comparator values ≤ 75 mg/dL and within ±20% of Comparator values \> 75 mg/dL).
Detailed Description: Design: Non-randomized, non-treatment, prospective, open label Study. After providing written informed consent, subjects meeting Inclusion/Exclusion Criteria will be connected to the System for a maximum of 72 hours. For the purposes of this clinical Study, the System will not display real-time glucose values, glucose trend graphs, or glucose alarms to either the Study Subjects or the Study Site Personnel. There will be no treatment of any Subject based on the output of the System. Diabetes management decisions throughout this Study (if applicable) will be conducted according to hospital procedures. * 1 set of laboratory measurements will be drawn per Subject for baseline. * A maximum of 82 blood samples will be drawn per Subject during the 72 hour monitoring session including. * Up to 75 blood samples drawn for comparative glucose measurement on laboratory analyzers. Blood samples will be taken from the sampling catheter in the peripheral vein in the contralateral extremity (relative to placement of the System) with no more than 25 samples taken within one day (00:00 to 23:59) for glucose measurement on laboratory analyzers (e.g., Yellow Springs Instrument glucose analyzer (YSI), Radiometer blood analyzer); * Up to three (3) sets of laboratory measurements; and * Up to 4 other blood samples, as needed. One week (7 + 3 days) after Sensor removal, subjects will be contacted for a follow-up assessment and document any subsequent adverse events.
Study: NCT02094859
Study Brief:
Protocol Section: NCT02094859