The Description Module contains narrative descriptions of the clinical trial, including a brief summary and detailed description. These descriptions provide important information about the study's purpose, methodology, and key details in language accessible to both researchers and the general public.

Description Module path is as follows:

Study -> Protocol Section -> Description Module

Description Module


Ignite Creation Date: 2025-12-25 @ 2:02 AM
Ignite Modification Date: 2025-12-25 @ 2:02 AM
NCT ID: NCT05530460
Brief Summary: The purpose of this study is to complete performance testing of our custom optical coherence tomography (OCT) device to verify it can deliver retinal images of similar quality to a commercial OCT device.
Detailed Description: There is some evidence that certain changes in retinal structures could be an indicator of Alzheimer's disease (AD). We are developing a new technology that combines optical coherence tomography (OCT) and angle-resolved low coherence interferometry (a/LCI) to make light scattering measurements of retinal tissue structure. The objective of this device feasibility study is to compare performance characteristics of our custom OCT instrument to a commercial OCT instrument. Once we have optimized our device, we may use it in future studies to identify retinal structure changes that could be possible AD biomarkers. For this study, the team will recruit healthy volunteers to attend one study visit that is about 30 minutes long. This visit will include examination with our custom OCT instrument, followed by examination with a commercial OCT system.