Viewing Study NCT03166059


Ignite Creation Date: 2025-12-24 @ 2:51 PM
Ignite Modification Date: 2026-02-21 @ 10:11 PM
Study NCT ID: NCT03166059
Status: UNKNOWN
Last Update Posted: 2017-12-22
First Post: 2017-05-22
Is NOT Gene Therapy: True
Has Adverse Events: False

Brief Title: CaveoVasc Thrombolysis Protection System, Access Protection Study
Sponsor: CaveoMed GmbH
Organization:

Study Overview

Official Title: Prospective, Single Arm, Study to Assess the Safety and Performance of the CaveoVasc® Thrombolysis Protection System, for Femoral Artery Access and Protection in Patients Treated With Thrombolysis for Limb Ischemia
Status: UNKNOWN
Status Verified Date: 2017-05
Last Known Status: NOT_YET_RECRUITING
Delayed Posting: No
If Stopped, Why?: Not Stopped
Has Expanded Access: False
If Expanded Access, NCT#: N/A
Has Expanded Access, NCT# Status: N/A
Acronym: None
Brief Summary: The CaveoVasc Thrombolysis Protection System is intended to aid sheath access and reduce bleeding complications during catheter directed thrombolysis (CDT) in the treatment of limb ischemia. This device has been designed to protect the puncture site by stabilizing the CDT sheath with the double-balloons that are part of the CaveoVasc system. Also the device is designed to reduce bleeding at the puncture site during CDT.

This study is designed to assess the safety and performance of the CaveoVasc Thrombolysis Protection System in twenty patients with limb ischemia undergoing treatment with CDT.
Detailed Description: Catheter directed thrombolysis (CDT) is a non-surgical option for many patients, which involves the insertion of an infusion catheter (via the femoral artery) into the thrombus allowing for slow delivery of a pharmacological thrombolytic agent. Insertion of the infusion catheter is performed in the catheterization laboratory. The patient is usually then transferred to the intensive care unit, with delivery catheter in place and the infusion running for at least 24 hours.

The major complications reported with CDT include major and minor bleeding, and complications secondary to early termination of the CDT due to access site bleeding.

CaveoMed developed the CaveoVasc® Thrombolysis Protection System, a vascular access protection device intended for use in thrombolysis procedures. Its function is to facilitate sheath access, and minimize risks of access site bleeding complications during lengthy catheter-directed thrombolysis procedures. Pressure balloons inflated outside the artery maintain a good seal throughout thrombolysis, which typically lasts up to 24 hours. The thrombolysis catheter is then removed, and hemostasis at the site is managed per the hospital's standard.

The CaveoVasc® Thrombolysis Protection System design has important advantages for use during CDT procedures (which require long catheter dwell times). The system is intended to protect from intra-procedural bleeding at the arterial access site, which currently remains a major problem for patients with critical limb ischemia undergoing CDT treatment. This protection would lower the rate of bleeding events, and increase the rate of completed intended therapy, without interruption of the CDT treatment due to bleeding complications.

Study Oversight

Has Oversight DMC: False
Is a FDA Regulated Drug?: False
Is a FDA Regulated Device?: False
Is an Unapproved Device?: None
Is a PPSD?: None
Is a US Export?: None
Is an FDA AA801 Violation?: