Viewing Study NCT03832205


Ignite Creation Date: 2025-12-25 @ 3:21 AM
Ignite Modification Date: 2026-03-02 @ 4:02 PM
Study NCT ID: NCT03832205
Status: COMPLETED
Last Update Posted: 2019-12-12
First Post: 2018-10-23
Is NOT Gene Therapy: True
Has Adverse Events: False

Brief Title: Validation of Respiratory Rate and Heart Rate Measurements by Capaciflectors Placed in Four Locations on the Chest
Sponsor: University Hospital Southampton NHS Foundation Trust
Organization:

Study Overview

Official Title: Validation of a Capaciflector Against a Pneumotachometer to Measure Respiratory Rate and a Capaciflector Against Electrocardiography to Measure Heart Rate at Different Anatomical Chest Locations During Cardiopulmonary Exercise Testing
Status: COMPLETED
Status Verified Date: 2019-12
Last Known Status: None
Delayed Posting: No
If Stopped, Why?: Not Stopped
Has Expanded Access: False
If Expanded Access, NCT#: N/A
Has Expanded Access, NCT# Status: N/A
Acronym: None
Brief Summary: In a collaborative project between University Hospital Southampton NHS Trust and the University of Southampton, the investigators have developed a wearable, mobile, non invasive, low-cost, continuous respiratory rate monitor device called a capaciflector.

This study will be the first ever study undertaken to attempt to measure the respiratory rate and heart rate on patients using a capaciflector. Investigators will evaluate whether the correlation between the capaciflector's measurements of respiratory rate and heart rate when compared with the gold standard pneumotachometer and electrocardiogram respectively is high enough to promote its potential future use within a device for clinical practice.
Detailed Description: RESEARCH QUESTION/AIM(S)

Will a capaciflector measure the respiratory rate as accurately as a pneumotachometer and the heart rate as accurately as an electrocardiogram?

Objectives:

1. To test the reliability and accuracy of our capaciflector in measuring respiratory rate and heart rate by comparing it with gold standard readings.
2. To assess whether different capaciflector locations around the body are all as accurate as each other.

STUDY DESIGN and METHODS of DATA COLLECTION AND DATA ANALYSIS

Unblinded observational study - for each subject demographic data will be collected: age, gender, height, weight, body mass index.

As well as the routine monitoring devices worn while undergoing a CPET test, the patients will have 4 capaciflectors placed on them as well. These will be placed using hypoallergenic medical grade tape.

* Over the left precordium
* Over the right precordium
* Under the left axilla
* Under the right axilla

Each practical component of the study setup will be delivered consistently by CI Dr Nick Hayward, Student Doctor Alan Doughty, or cardiopulmonary exercise testing (CPET) technicians once the study becomes established. The patient will then undergo a routine CPET and our monitors will not affect what a patient is asked to do as part of the current CPET protocol. If a patient is unable to finish the CPET then their data will still be included.

Raw CPET data will be exported via a feed from the CPET machine and also from the capaciflectors. A laptop computer provided by the University will receive both data feeds. Simultaneous recordings will be taken from each of the capaciflectors at time zero and then at one minute intervals. These anonymous datasets will be stored on the laptop computer. Once the CPET test is completed, the participant's involvement in the study will have ended.

Anonymised raw data analysis comparisons will be made by the Department of Engineering, University of Southampton, in collaboration with the other investigators of this study. There will be no patient identifiable information in any dataset, yet demographic data and raw CPET acquired data will be held securely on the laptop in line with clinical governance requirements of University Hospital Southampton. Statistical support by Dr Helen Moyses of the Clinical Informatics Research Unit, Southampton, will be provided for data analysis at the final stage.

STUDY SETTING

The study will be a single centre study at University Hospital Southampton. Participants will have been recruited from the patients booked for a CPET test prior to them having major surgery. The study will be conducted in the CPET department in the Day Surgery Unit. This is where the CPET for patients routinely takes place.

Study Oversight

Has Oversight DMC: False
Is a FDA Regulated Drug?: False
Is a FDA Regulated Device?: False
Is an Unapproved Device?: None
Is a PPSD?: None
Is a US Export?: None
Is an FDA AA801 Violation?: